Q Exactive Hybrid Quadrupole-Orbitrap Mass Spectrometer (Q Exactive Plus)

Brand: Thermo Scientific 
Model: Q Exactive Plus 
Fixed Asset Number: S210001965 
Affiliated Core: Multi-omics Mass Spectrometry Core
Location: Room 116, Building B, GaoKe Innovation Center
Equipment Manager: Zhao Gongzheng, Zheng Ting, Li E
Contact: 19129190970, zhaogz@szbl.ac.cn

1. The mass analyzer combines a quadrupole with an electrostatic field orbitrap, offering a mass range of 50-6000 m/z.

2. The instrument has a resolution of 140,000 FWHM (for m/z≤200) and features ≥4 adjustable settings.
3. The high-resolution mass spectrometry acquisition rate can reach up to 12 Hz.
4. Stability: After calibration, without re-calibration within 48 hours, repeated injections of 100 fg reserpine result in a mass accuracy of ≤2 ppm at 609 m/z.
5. Positive/negative ion switching speed is less than 1 s
6. Sensitivity: Full Scan sensitivity (with resolution ≥70000 FWHM) (m/z 100-900), 50 fg reserpine injection achieves S/N>500:1; SIM sensitivity (with resolution ≥70000 FWHM): 50 fg reserpine injection achieves S/N>1000:1.
7. Detector: FT intact detection.