Brand:Agilent technologies
Model:8890-7010B
Fixed Asset Number:S210002154
Affiliated Core:Multi-omics Mass Spectrometry Core
Location:115,Building B, Gaoke Innovation Center
Equipment Manager:Zhao Gongzheng, Zheng Ting, Li E
Contact:19129910970,zhaogz@szbl.ac.cn
1. Mass Range: 10–1050 m/z
2. Resolution: Adjustable from 0.4 to 4 amu
3. Detection Limit & Sensitivity:
-
-
IDL (Instrument Detection Limit): ≤0.5 fg OFN (octafluoronaphthalene), based on 8 consecutive injections of 2 fg
-
PCI-MRM Sensitivity: 100 fg BZP (benzophenone), S/N ≥50:1 (transition m/z 183→105)
-
4. Collision Cell Helium Quench: Eliminates background noise from helium carrier gas; Quench gas flow adjustable from 0 to 5.0 mL/min
5. MRM Scan Rate:
-
Maximum: ≥800 MRM transitions/second
-
Minimum SRM dwell time: 0.5 ms
6. Ion Source:
-
Electron Ionization (EI) source
-
Independently temperature-controlled
-
Maximum temperature: 350°C
7. Mass Analyzer:
-
Gold-coated quartz hyperbolic quadrupoles
-
No pre-filter quadrupole
-
Independent temperature control
-
Maximum temperature: 200°C
-
-
-
-